Title: Evaluation of Polyurethane Nasolacrimal Duct Stents: In Vivo Studies in New Zealand Rabbits

Journal Article · · Cardiovascular and Interventional Radiology
;  [1];  [2];  [1];  [3]
  1. University Hospital Bonn, Department of Radiology (Germany)
  2. Clinics of the University of Bonn, Haus fuer Experimentelle Therapie (Germany)
  3. Martin Luther University of Halle-Wittenberg, Department of Anatomy and Cell Biology (Germany)

The purpose of this study was to evaluate the radiographic and biological effects of different polyurethane nasolacrimal duct stents in an animal model. Fifteen polyurethane nasolacrimal duct stents (n = 5 mushroom-type stents, n = 5 newly designed S-shaped TearLeader stents without hydrophilic coating, and n = 5 S-shaped TearLeader stents with hydrophilic coating) were implanted in the nasolacrimal ducts of eight unaffected New Zealand rabbits. One nasolacrimal system served as control. Clinical and radiographic follow-up was performed at 1-, 2-, and 4-week intervals, then after a 3-month interval, after which the animals were euthanized. All stents were implanted without major periprocedural complications. The stents proved to be patent by the end of the procedure. During follow-up, all mushroom-type stents were occluded at 4 weeks. None of these stents opened to forced irrigation. Clinically, all rabbits demonstrated severe dacryocystitis. Three out of five TearLeader stents without hydrophilic coating were blocked at 4 weeks; one out of five was open to irrigation. Best results were observed in the stent group with hydrophilic coating. Follow-up dacryocystography demonstrated patent stents in nasolacrimal ducts of all animals after 4 weeks. In only one of five cases, the coated stent became partially occluded after 2 months. These animals were free of clinical symptoms. After 3 months, at least three out of five stents still opened to forced irrigation and only one stent was completely blocked. Dislocation of the stents was not observed. Refinement of the stent surface and stent design improves the results of nasolacrimal duct stenting in this animal model. Implantation of hydrophilic-coated S-shaped stents is highly superior to conventional mushroom-type stents and noncoated stent types. Hydrophilic coating seems to prevent foreign-body reactions, resulting in maximized stent patency.

OSTI ID:
21091137
Journal Information:
Cardiovascular and Interventional Radiology, Vol. 29, Issue 5; Other Information: DOI: 10.1007/s00270-005-0232-8; Copyright (c) 2006 Springer Science+Business Media, Inc.; www.springer-ny.com; Country of input: International Atomic Energy Agency (IAEA); ISSN 0174-1551
Country of Publication:
United States
Language:
English

Similar Records

Nasolacrimal Polyurethane Stent: Complications with CT Correlation
Journal Article · Sun Nov 15 00:00:00 EST 1998 · Cardiovascular and Interventional Radiology · OSTI ID:21091137

Nasolacrimal Stenting: Toward Improving Outcomes with a Simple Modification of the Song Stent
Journal Article · Tue Aug 15 00:00:00 EDT 2006 · Cardiovascular and Interventional Radiology · OSTI ID:21091137

Treatment of Complete and Partial Obstruction of the Nasolacrimal System with Polyurethane Stents: Initial Experience
Journal Article · Thu Jan 15 00:00:00 EST 1998 · Cardiovascular and Interventional Radiology · OSTI ID:21091137