Successful Transcatheter Embolotherapy with a New Platinum Microcoil: The Berenstein Liquid Coil
- Department of Radiology, Kansai Medical University, 10-15 Fumizono-cho, Moriguchi-city, Osaka, 570 (Japan)
Purpose: To determine the usefulness of a new platinum microcoil, the Berenstein Liquid Coil for vascular embolization. Methods: Nine patients underwent transcatheter arterial embolization with liquid coils. The occluded vessels included the splenic artery in four patients, gastroduodenal artery in two, arteriovenous malformation or fistula in two, and middle hepatic artery in one. All coils were delivered with a saline flush through a Tracker-18 catheter. Results: All arteries except one were successfully occluded in a one-stage procedure using liquid coils. No friction was observed between the liquid coils and the microcatheter regardless of the tortuosity of the artery. Platelet counts in all patients had increased after splenic embolization. In two patients with liver tumors, perfusion to the inappropriate area during intraarterial chemotherapy was prevented by selective embolization. Life-threatening bleeding was successfully treated by gastroduodenal artery occlusion. No complications were encountered. Conclusion: The Berenstein Liquid Coil provides easy, safe, and rapid vascular occlusion.
- OSTI ID:
- 21080386
- Journal Information:
- Cardiovascular and Interventional Radiology, Vol. 21, Issue 4; Other Information: DOI: 10.1007/s002709900264; Copyright (c) 1996 Springer-Verlag New York Inc; Country of input: International Atomic Energy Agency (IAEA); ISSN 0174-1551
- Country of Publication:
- United States
- Language:
- English
Similar Records
Renal Arteriovenous Fistula with Rapid Blood Flow Successfully Treated by Transcatheter Arterial Embolization: Application of Interlocking Detachable Coil as Coil Anchor
Angiographic Findings and Embolotherapy in Renal Arterial Trauma