Title: A method for synchronizing an external respiratory signal with a list-mode PET acquisition

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.2791036· OSTI ID:21032856

A method is proposed to synchronize positron emission tomography (PET) list-mode data with an externally recorded respiratory signal in the absence of a master clock. When the respiratory signal reaches a user-defined threshold, a trigger mark is stored in the list-mode file. After the acquisition, synchronization is achieved when the stored trigger marks are superimposed on the respiratory curve to form a horizontal line over time at the user-defined threshold. Synchronization was possible and unequivocal for ten out of ten clinical studies. The list-mode acquisition actually started approximately 40 and 4 s after acquisition initiation at the user interface of the Philips Gemini and the GE DLS PET-CT systems, respectively.

OSTI ID:
21032856
Journal Information:
Medical Physics, Vol. 34, Issue 11; Other Information: DOI: 10.1118/1.2791036; (c) 2007 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States