Slag material produced by the Plasma Hearth Process (PHP) varies in chemical composition due to the heterogeneous nature of the input sample feed. X-ray fluorescence (XRF) is a spectroscopic technique which has been evaluated to perform elemental analyses on surrogate slag material for process control. The intensity of Si, Al and Fe in the slag samples was utilized to determine the appropriate matrix standard set for the determination of Ce. The precision of the XRF technique was better than 5% RSD. The limit of detection for Ce varied with sample matrix and was typically below 0.01 % by weight. The linear dynamic range for the technique was evaluated over 2 orders of magnitude. The Ce determinations performed directly on slag material by the XRF technique were similar to ICP-AES analyses. No addition waste streams were created from the analyses by the XRF technique.
- OSTI ID:
- 210230
- Report Number(s):
- CONF-9509139-; TRN: 95:008324-0071
- Resource Relation:
- Conference: 7. ACS special symposium: emerging technologies in hazardous waste management, Atlanta, GA (United States), 17-20 Sep 1995; Other Information: PBD: 1995; Related Information: Is Part Of Emerging technologies in hazardous waste management VII; Tedder, D.W. [ed.]; PB: 1352 p.
- Country of Publication:
- United States
- Language:
- English
Similar Records
Surrogate formulations for thermal treatment of low-level mixed waste. Part 3: Plasma hearth process testing
Results of bench-scale plasma system testing in support of the Plasma Hearth Process