High-energy collision induced dissociation fragmentation pathways of peptides, probed using a multiturn tandem time-of-flight mass spectrometer 'MULTUM-TOF/TOF'
- Department of Physics, Graduate School of Science, Osaka University, 1-1 Machikaneyama, Toyonaka, Osaka 560-0043 (Japan)
A new multiturn tandem time-of-flight (TOF) mass spectrometer 'MULTUM-TOF/TOF' has been designed and constructed. It consists of a matrix-assisted laser desorption/ionization ion source, a multiturn TOF mass spectrometer, a collision cell, and a quadratic-field ion mirror. The multiturn TOF mass spectrometer can overcome the problem of precursor ion selection in TOF, due to insufficient time separation between two adjacent TOF peaks, by increasing the number of cycles. As a result, the total TOF increases with the increase in resolving power. The quadratic-field ion mirror allows temporal focusing for fragment ions with different kinetic energies. Product ion spectra from monoisotopically selected precursor ions of angiotensin I, substance P, and bradykinin have been obtained. The fragment ions observed are mainly the result of high-energy collision induced dissociation.
- OSTI ID:
- 20953532
- Journal Information:
- Review of Scientific Instruments, Vol. 78, Issue 7; Other Information: DOI: 10.1063/1.2751403; (c) 2007 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
- Country of Publication:
- United States
- Language:
- English
Similar Records
Mechanisms of Peptide Fragmentation from Time-and Energy-Resolved Surface-Induced Dissociation Studies: Dissociation of Angiotensin Analogs
Tandem time-of-flight mass spectrometer for cluster--surface scattering experiments
Related Subjects
37 INORGANIC
ORGANIC
PHYSICAL AND ANALYTICAL CHEMISTRY
ANGIOTENSIN
BRADYKININ
CHEMICAL ANALYSIS
COLLISIONS
DESORPTION
DISSOCIATION
FOCUSING
FRAGMENTATION
ION SOURCES
IONIZATION
IONS
KINETIC ENERGY
MIRRORS
PRECURSOR
TIME-OF-FLIGHT MASS SPECTROMETERS