Design and construction of uniform glow discharge plasma system operating under atmospheric condition
Journal Article
·
· Review of Scientific Instruments
- Biomedical Engineering Department, Baskent University, Ankara 06530 (Turkey)
The design of a uniform glow discharge plasma system operating without vacuum is presented. A full-bridge switching circuit was used to switch the transformers. The primary windings of transformers were connected in parallel, but in opposite phase to double the output voltage. Theoretically, 20 000 V{sub pp} was obtained. Rectangle copper electrodes were used, and placed parallel to each other. To prevent the spark production that is, to obtain uniformity, two 2 mm Teflon sheets were glued to the electrodes. However, it was observed that the operating frequency also affected the uniformity. For the system presented here, the frequency at which more uniformity was obtained was found to be 14 kHz.
- OSTI ID:
- 20953465
- Journal Information:
- Review of Scientific Instruments, Vol. 78, Issue 6; Other Information: DOI: 10.1063/1.2745230; (c) 2007 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
- Country of Publication:
- United States
- Language:
- English
Similar Records
The One Atmosphere Uniform Glow Discharge Plasma (OAUGDP) as a classical normal glow discharge
Industrial applications of the one atmosphere uniform glow discharge plasma
A polyphase low frequency RF power supply for a One Atmosphere Uniform Glow Discharge Plasma (OAUGDP)
Conference
·
Thu Jul 01 00:00:00 EDT 1999
·
OSTI ID:20953465
+2 more
Industrial applications of the one atmosphere uniform glow discharge plasma
Conference
·
Thu Dec 31 00:00:00 EST 1998
·
OSTI ID:20953465
A polyphase low frequency RF power supply for a One Atmosphere Uniform Glow Discharge Plasma (OAUGDP)
Conference
·
Thu Jul 01 00:00:00 EDT 1999
·
OSTI ID:20953465