Compact radiometric microwave calibrator
Journal Article
·
· Review of Scientific Instruments
- SSAI Goddard Spaceflight Center, Greenbelt, Maryland 20771 (United States)
The calibration methods for the ARCADE II instrument are described and the accuracy estimated. The Steelcast coated aluminum cones which comprise the calibrator have a low reflection while maintaining 94% of the absorber volume within 5 mK of the base temperature (modeled). The calibrator demonstrates an absorber with the active part less than one wavelength thick and only marginally larger than the mouth of the largest horn and yet black (less than -40 dB or 0.01% reflection) over five octaves in frequency.
- OSTI ID:
- 20853300
- Journal Information:
- Review of Scientific Instruments, Vol. 77, Issue 6; Other Information: DOI: 10.1063/1.2209960; (c) 2006 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
- Country of Publication:
- United States
- Language:
- English
Similar Records
A Liquid-Helium-Cooled Absolute Reference Cold Load forLong-Wavelength Radiometric Calibration
A liquid-helium-cooled absolute reference cold load for long-wavelength radiometric calibration
Radiometric sources for the Los Alamos National Laboratory calibration Laboratory
Journal Article
·
Tue May 01 00:00:00 EDT 1990
· Review of Scientific Instruments
·
OSTI ID:20853300
+3 more
A liquid-helium-cooled absolute reference cold load for long-wavelength radiometric calibration
Journal Article
·
Thu Oct 01 00:00:00 EDT 1992
· Review of Scientific Instruments; (United States)
·
OSTI ID:20853300
+3 more
Radiometric sources for the Los Alamos National Laboratory calibration Laboratory
Conference
·
Fri Jul 01 00:00:00 EDT 1994
·
OSTI ID:20853300
+4 more