skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Dose distribution from x-ray microbeam arrays applied to radiation therapy: An EGS4 Monte Carlo study

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.1951043· OSTI ID:20726259
; ; ; ; ;  [1]
  1. INFM-OGG, c/o ESRF, BP220, F-38043 Grenoble Cedex (France)

We present EGS4 Monte Carlo calculations of the spatial distribution of the dose deposited by a single x-ray pencil beam, a planar microbeam, and an array of parallel planar microbeams as used in radiation therapy research. The profiles of the absorbed dose distribution in a phantom, including the peak-to-valley ratio of the dose distribution from microbeam arrays, were calculated at micrometer resolution. We determined the dependence of the findings on the main parameters of photon and electron transport. The results illustrate the dependence of the electron range and the deposited in-beam dose on the cut-off energy, of the electron transport, as well as the effects on the dose profiles of the beam energy, the array size, and the beam spacing. The effect of beam polarization also was studied for a single pencil beam and for an array of parallel planar microbeams. The results show that although the polarization effect on the dose distribution from a 3 cmx3 cm microbeam array inside a water phantom is large enough to be measured at the outer side of the array (16% difference of the deposited dose for x-ray beams of 200 keV), it is not detectable at the array's center, thus being irrelevant for the radiation therapy purposes. Finally we show that to properly compare the dose profiles determined with a metal oxide semiconductor field emission transistor detector with the computational method predictions, it is important to simulate adequately the size and the material of the device's Si active element.

OSTI ID:
20726259
Journal Information:
Medical Physics, Vol. 32, Issue 8; Other Information: DOI: 10.1118/1.1951043; (c) 2005 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English