skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Cross calibration of new x-ray films against direct exposure film from 1 to 8 keV using the X-pinch x-ray source

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.2135276· OSTI ID:20723239
; ; ; ; ;  [1]
  1. Laboratory of Plasma Studies, Cornell University, Ithaca, New York 14853 (United States)

A cross calibration of readily available x-ray sensitive films has been carried out against the calibrated direct exposure film (DEF) which is no longer being manufactured by Kodak. Four-wire X pinches made from various metal wires were used as x-ray sources for this purpose. Tests were carried out for the Kodak films Biomax MS, Biomax XAR, M100, Technical Pan, and T-Max over the energy range of 1-8 keV (12.4-1.5 A wavelength). The same hand-development procedures as described by Henke et al. [J. Opt. Soc. Am. B 3, 1540 (1986)] were followed for all films in every test. Sensitivity curves as a function of wavelength for these films relative DEF are presented. These relative calibrations show that Biomax MS is likely to be the best replacement film for DEF for most purposes over the energy range tested here.

OSTI ID:
20723239
Journal Information:
Review of Scientific Instruments, Vol. 76, Issue 11; Other Information: DOI: 10.1063/1.2135276; (c) 2005 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English

Similar Records

Characterization and cross calibration of Agfa D4, D7, and D8 and Kodak SR45 x-ray films against direct exposure film at 4.0-5.5 keV
Journal Article · Sat Apr 15 00:00:00 EDT 2006 · Review of Scientific Instruments · OSTI ID:20723239

Absolute calibration of Kodak Biomax-MS film response to x rays in the 1.5- to 8-keV energy range
Journal Article · Sun Oct 15 00:00:00 EDT 2006 · Review of Scientific Instruments · OSTI ID:20723239

Response Model for Kodak Biomax-MS Film to X Rays
Journal Article · Wed Jan 24 00:00:00 EST 2007 · Review of Scientific Instruments · OSTI ID:20723239