Title: Comparison of Indiana University Cyclotron Facility Faraday cup proton dosimetry with radiochromic films, a calorimeter, and a calibrated ion chamber

Journal Article · · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers)
DOI:https://doi.org/10.1109/23.819151· OSTI ID:20014738

The accuracy and utility of the dosimetry system used for radiation effects research with high energy protons at the Indiana University Cyclotron Facility, IUCF, has been confirmed by comparison with an independently calibrated Markus ion chamber, a Schulz water calorimeter and GAFCHROMIC{trademark} films.