Design of a platinum resistance thermometer temperature measuring transducer and improved accuracy of linearizing the output voltage
An improved method is presented for designing a temperature measuring transducer, the electrical circuit of which comprises an unbalanced bridge, in one arm of which is a platinum resistance thermometer, and containing a differential amplifier with feedback. Values are given for the coefficients, the minimum linearization error is determined, and an example is also given of the practical design of the transducer, using the given coefficients. A determination is made of the limiting achievable accuracy in linearizing the output voltage of the measuring transducer, as a function of the range of measured temperature.
- Sponsoring Organization:
- USDOE
- OSTI ID:
- 183617
- Journal Information:
- Measurement Techniques, Vol. 38, Issue 1; Other Information: PBD: Jun 1995; TN: Translated from Izmeritel`naya Tekhnika; No. 1, 48-50(Jan 1995)
- Country of Publication:
- United States
- Language:
- English
Similar Records
Errors of instrument transducers of temperature to unified dc voltage
Dynamic response of a feedback thermoresistive electrical substitution pyranometer
Dynamic response of an electronic feedback thermoresistive electrical substitution pyranometer
Journal Article
·
Mon Apr 01 00:00:00 EST 1985
· Meas. Tech. (Engl. Transl.); (United States)
·
OSTI ID:183617
Dynamic response of a feedback thermoresistive electrical substitution pyranometer
Journal Article
·
Fri May 01 00:00:00 EDT 1998
· Journal of Solar Energy Engineering
·
OSTI ID:183617
+3 more
Dynamic response of an electronic feedback thermoresistive electrical substitution pyranometer
Conference
·
Wed Nov 01 00:00:00 EST 1995
·
OSTI ID:183617
+2 more