Title: Noise and resolution of Bayesian reconstruction for multiple image configurations

Journal Article · · IEEE Transactions on Nuclear Science
DOI:https://doi.org/10.1109/23.273561· OSTI ID:142401
;  [1]
  1. Univ. of California, Los Angeles, CA (United States)

Images reconstructed by Bayesian and maximum-likelihood (ML) using a Gibbs prior with prior weight {beta} were compared with images produced by filtered back projection (FBP) from sinogram data simulated with different counts and image configurations. Bayesian images were generated by the OSL algorithm accelerated by an over relaxation parameter. For relatively low {beta}, Bayesian images can yield an overall improvement to the images compared to ML reconstruction. However, for larger {beta}, Bayesian images degrade from the standpoint of noise and quantitation. Compared to FBP, the ML images were superior in a mean square error sense in regions of low activity level and for small structures. At a comparable noise level to FBP, Bayesian reconstruction can be used to effectively recover higher resolution images. The overall performance is dependent on the image structure and the weight of the Bayesian prior.

DOE Contract Number:
FC03-87ER60615
OSTI ID:
142401
Journal Information:
IEEE Transactions on Nuclear Science, Vol. 40, Issue 6; Other Information: PBD: Dec 1993
Country of Publication:
United States
Language:
English

Similar Records

Assessment of volumetric noise and resolution performance for linear and nonlinear CT reconstruction methods
Journal Article · Tue Jul 15 00:00:00 EDT 2014 · Medical Physics · OSTI ID:142401

Precision and accuracy of regional radioactivity quantitation using the maximum likelihood EM reconstruction algorithm
Journal Article · Thu Sep 01 00:00:00 EDT 1994 · IEEE Transactions on Medical Imaging (Institute of Electrical and Electronics Engineers); (United States) · OSTI ID:142401

An investigation of filter choice for filtered back-projection reconstruction in PET
Journal Article · Mon Jun 01 00:00:00 EDT 1998 · IEEE Transactions on Nuclear Science · OSTI ID:142401