Approximate reconstruction of attenuation map in SPECT imaging
- Russian Academy of Sciences, Novosibirsk (Russian Federation). Inst. for Theoretical and Applied Mechanics
The influence of linear attenuation presents certain difficulties for quantitative single-photon emission computed tomography (SPECT) in medical imaging. To apply an attenuation correction algorithm, one has first to estimate the attenuation map from some additional measurements or by the use of SPECT data. The determination of the attenuation map from SPECT-type data may also be considered as a special problem of nondestructive evaluation of materials. A new technique for obtaining the attenuation map from SPECT data is proposed. Unlike other methods, the approach suggested demands no a priori knowledge of the problem. The method has been derived on the basis of the approximate linear relation between SPECT data and the attenuation map. The approach permits local changes of the attenuation coefficient to be reconstructed rather than its actual value. In this way, such details of the attenuation map as spots, curves and edges can be localized. Limitations of application of the suggested technique to real data are considered. Results of numerical evaluation and real data processing are presented.
- OSTI ID:
- 131647
- Journal Information:
- IEEE Transactions on Nuclear Science, Vol. 42, Issue 5; Other Information: PBD: Oct 1995
- Country of Publication:
- United States