Effective efficiency measurement results conducted on a x-band thin film waveguide barretter mount developed at NPL, New Delhi
Conference
·
OSTI ID:125876
- National Physical Lab., New Delhi (India)
Effective efficiency measurements have been conducted at 10.0 GHz using a single load microcalorimetric technique on a thin film primary standard waveguide barretter mount designed and fabricated at National Physical Laboratory, New Delhi. The same mount was taken to three National Laboratories, namely NIST, USA; KRISS, Republic of Korea; PTB, Germany for an international comparison of effective efficiency measurements. This paper briefly describes the techniques employed for the measurement of effective efficiency of bolometer mounts by these countries. The results of effective efficiency are reported and intercomared.
- OSTI ID:
- 125876
- Report Number(s):
- CONF-940603-; TRN: 95:005769-0268
- Resource Relation:
- Conference: CPEM `94: conference on precision electromagnetic measurements, Boulder, CO (United States), 27 Jun - 1 Jul 1994; Other Information: PBD: 1994; Related Information: Is Part Of 1994 Conference on precision electromagnetic measurements digest; DeWeese, E.; Bennett, G. [eds.]; PB: 577 p.
- Country of Publication:
- United States
- Language:
- English
Similar Records
Application of laser induced breakdown spectroscopy (LIBS) instrumentation for international safeguards
Partnerships for Clean Development and Climate: Business andTechnology Cooperation Benefits
Calibration and Characterization of the Small Sample Calorimeter
Conference
·
Fri Jan 01 00:00:00 EST 2010
·
OSTI ID:125876
+3 more
Partnerships for Clean Development and Climate: Business andTechnology Cooperation Benefits
Conference
·
Tue Aug 22 00:00:00 EDT 2006
·
OSTI ID:125876
+2 more
Calibration and Characterization of the Small Sample Calorimeter
Technical Report
·
Mon Aug 13 00:00:00 EDT 2012
·
OSTI ID:125876