skip to main content

SciTech ConnectSciTech Connect

Title: Timing Characteristics of Large Area Picosecond Photodetectors

The LAPPD Collaboration was formed to develop ultralast large-area imaging photodetectors based on new methods for fabricating microchannel plates (MCPs). In this paper we characterize the time response using a pulsed, sub picosecond laser. We observe single photoelectron time resolutions of a 20 cm x 20 cm MCP consistently below 70 ps, spatial resolutions of roughly 500 pm, and median gains higher than 10(7). The RMS measured at one particular point on an LAPPD detector is 58 ps, with in of 47 ps. The differential time resolution between the signal reaching the two ends of the delay line anode is measured to be 5.1 ps for large signals, with an asymptotic limit falling below 2 ps as noise-over-signal approaches zero.
; ; ; ; ; ; ; ; ;
Publication Date:
OSTI Identifier:
DOE Contract Number:
Resource Type:
Journal Article
Resource Relation:
Journal Name: Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment; Journal Volume: 795
Research Org:
Argonne National Laboratory (ANL)
Sponsoring Org:
USDOE Office of Science - Office of Basic Energy Sciences; USDOE Office of Science - Office of High Energy Physics; National Science Foundation (NSF)
Country of Publication:
United States
LAPPD; MCP; Single photon; TOF; TTS; large area; photodetector; picosecond; resolution