Title: Validation and Verification of the Correct Operation of the Thermo Fisher Scientific’s Genesys 10S spectrometer used in ANC122 and NF-ANC122

In December of 2015, the old qualified Prim UV-Vis instrument used in the ANC122 and NFANC122 procedures failed its monthly calibration check. Attempts to get the system to pass by cleaning flow cells, replacing the light bulb and realigning the light path failed to get sustained reproducible results from the system. As it could no longer pass QA/QC requirements the decision to take it out of service was made. To replace the system, a previously purchased Thermo Fisher Scientific UV-VIS spectrometer (model Genesys 10S; Serial #: 2L5R059137) was retrieved from storage. This report shows that the system is performing as required and meeting all QC requirements listed in ANC122 and NF-ANC122