skip to main content

SciTech ConnectSciTech Connect

Title: Calibration and data collection protocols for reliable lattice parameter values in electron pair distribution function studies

Different protocols for calibrating electron pair distribution function (ePDF) measurements are explored and described for quantitative studies on nanomaterials. It is found that the most accurate approach to determine the camera length is to use a standard calibration sample of Au nanoparticles from the National Institute of Standards and Technology. Different protocols for data collection are also explored, as are possible operational errors, to find the best approaches for accurate data collection for quantitative ePDF studies.
 [1] ;  [1] ;  [1] ;  [1] ;  [2]
  1. Brookhaven National Lab. (BNL), Upton, NY (United States)
  2. Brookhaven National Lab. (BNL), Upton, NY (United States); Columbia Univ., New York, NY (United States)
Publication Date:
OSTI Identifier:
Report Number(s):
Journal ID: ISSN 1600-5767; JACGAR
DOE Contract Number:
Resource Type:
Journal Article
Resource Relation:
Journal Name: Journal of Applied Crystallography (Online); Journal Volume: 48; Journal Issue: 1
International Union of Crystallography
Research Org:
Brookhaven National Laboratory (BNL), Upton, NY (United States)
Sponsoring Org:
USDOE Office of Science (SC), Basic Energy Sciences (BES) (SC-22)
Country of Publication:
United States
77 NANOSCIENCE AND NANOTECHNOLOGY; 46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY electron pair distribution function; calibration; data collection; nanmaterials