Ion Trapping, Storage, and Ejection in Structures for Lossless Ion Manipulations
- Pacific Northwest National Lab. (PNNL), Richland, WA (United States). Biological Sciences Division
Here, a structure for lossless ion manipulation (SLIM) module was constructed with electrode arrays patterned on a pair of parallel printed circuit boards (PCB) separated by 5 mm and utilized to investigate capabilities for ion trapping at 4 Torr. Positive ions were confined by application of RF having alternating phases on a series of inner rung electrodes and by positive DC potentials on surrounding guard electrodes on each PCB. An axial DC field was also introduced by stepwise varying the DC potential of the inner rung electrodes so as to control the ion transport and accumulation inside the ion trap. We show that ions could be trapped and accumulated with 100% efficiency, stored for at least 5 hours with no losses, and could be rapidly ejected from the SLIM trap.
- Research Organization:
- Pacific Northwest National Laboratory (PNNL), Richland, WA (United States). Environmental Molecular Sciences Laboratory (EMSL)
- Sponsoring Organization:
- USDOE Office of Science (SC), Biological and Environmental Research (BER); USDOE Laboratory Directed Research and Development (LDRD) Program
- Grant/Contract Number:
- AC05-76RL01830
- OSTI ID:
- 1208779
- Report Number(s):
- PNNL-SA-107651; 48680; 48135; KP1601010
- Journal Information:
- Analytical Chemistry, Vol. 87, Issue 12; ISSN 0003-2700
- Publisher:
- American Chemical Society (ACS)Copyright Statement
- Country of Publication:
- United States
- Language:
- English
Web of Science
Ultra-high-resolution ion mobility spectrometry—current instrumentation, limitations, and future developments
|
journal | April 2019 |
Recent technical advances in proteomics
|
journal | January 2019 |
Advances in targeted proteomics and applications to biomedical research
|
journal | August 2016 |
Nanowell-mediated multidimensional separations combining nanoLC with SLIM IM-MS for rapid, high-peak-capacity proteomic analyses
|
journal | November 2018 |
Profiling microbial lignocellulose degradation and utilization by emergent omics technologies
|
journal | July 2016 |
Similar Records
Experimental Evaluation and Optimization of Structures for Lossless Ion Manipulations for Ion Mobility Spectrometry with Time-of-Flight Mass Spectrometry
Rectangular Ion Funnel: A New Ion Funnel Interface for Structures for Lossless Ion Manipulations