Liquid scintillation counters measure charged particle-emitting radioactive isotopes and are used for environmental studies, nuclear chemistry, and life science. Alpha and beta emissions arising from the material under study interact with the scintillation cocktail to produce light. The prototypical liquid scintillation counter employs low-level photon-counting detectors to measure the arrival of the scintillation light produced as a result of the dissolved material under study interacting with the scintillation cocktail. For reliable operation the counting instrument must convey the scintillation light to the detectors efficiently and predictably. Current best practices employ the use of two or more detectors for coincidence processing to discriminate true scintillation events from background events due to instrumental effects such as photomultiplier tube dark rates, tube flashing, or other light emission not generated in the scintillation cocktail vial. In low background liquid scintillation counters additional attention is paid to shielding the scintillation cocktail from naturally occurring radioactive material (NORM) present in the laboratory and within the instruments construction materials. Low background design is generally at odds with optimal light collection. This study presents the evolution of a light collection design for liquid scintillation counting in a low background shield. The basic approach to achieve both good light collection and a low background measurement is described. The baseline signals arising from the scintillation vial are modeled and methods to efficiently collect scintillation light are presented as part of the development of a customized low-background, high sensitivity liquid scintillation counting system.

Research Organization:
Pacific Northwest National Lab. (PNNL), Richland, WA (United States)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC05-76RL01830
OSTI ID:
1182862
Report Number(s):
PNNL-SA-104952
Journal Information:
Applied Optics, 54(9):2413-2423, Journal Name: Applied Optics, 54(9):2413-2423
Country of Publication:
United States
Language:
English

Similar Records

A PROTO-TYPE LIQUID SCINTILLATION COUNTER USING CURRENT MODE IMPEDANCE MATCHING FOR LOW ENERGY BETA COUNTING
Journal Article · Tue Jan 01 00:00:00 EST 1963 · Proc. Japan Conf. Radioisotopes · OSTI ID:1182862

Low Level Calibration of Liquid Scintillation Counting
Conference · Wed Jul 01 00:00:00 EDT 2020 · OSTI ID:1182862

A DEVICE TO MEASURE LOW LEVELS OF RADIOACTIVE CONTAMINANTS IN ULTRA-CLEAN MATERIALS
Technical Report · Fri Mar 17 00:00:00 EST 2006 · OSTI ID:1182862