Title: Test Report for NG Sensors GTX-1000.

This report describes initial testing of the NG Sensor GTX-1000 natural gas monitoring system. This testing showed that the retention time, peak area stability and heating value repeatability of the GTX-1000 were promising for natural gas measurements in the field or at the well head. The repeatability can be less than 0.25% for LHV and HHV for the Airgas standard tested in this report, which is very promising for a first generation prototype. Ultimately this system should be capable of 0.1% repeatability in heating value at significant size and power reductions compared with competing systems.