Title: Automated High-Pressure Titration System with In Situ Infrared Spectroscopic Detection

Journal Article · · Review of Scientific Instruments, 85(4):Article No. 044102
DOI:https://doi.org/10.1063/1.4870411· OSTI ID:1159012

A fully automated titration system with infrared detection was developed for investigating interfacial chemistry at high pressures. The apparatus consists of a high-pressure fluid generation and delivery system coupled to a high-pressure cell with infrared optics. A manifold of electronically actuated valves is used to direct pressurized fluids into the cell. Precise reagent additions to the pressurized cell are made with calibrated tubing loops that are filled with reagent and placed in-line with the cell and a syringe pump. The cell’s infrared optics facilitate both transmission and attenuated total reflection (ATR) measurements to monitor bulk-fluid composition and solid-surface phenomena such as adsorption, desorption, complexation, dissolution, and precipitation. Switching between the two measurement modes is accomplished with moveable mirrors that direct radiation from a Fourier transform infrared spectrometer into the cell along transmission or ATR light paths. The versatility of the high-pressure IR titration system is demonstrated with three case studies. First, we titrated water into supercritical CO2 (scCO2) to generate an infrared calibration curve and determine the solubility of water in CO2 at 50 °C and 90 bar. Next, we characterized the partitioning of water between a montmorillonite clay and scCO2 at 50 °C and 90 bar. Transmission-mode spectra were used to quantify changes in the clay’s sorbed water concentration as a function of scCO2 hydration, and ATR measurements provided insights into competitive residency of water and CO2 on the clay surface and in the interlayer. Finally, we demonstrated how time-dependent studies can be conducted with the system by monitoring the carbonation reaction of forsterite (Mg2SiO4) in water-bearing scCO2 at 50 °C and 90 bar. Immediately after water dissolved in the scCO2, a thin film of adsorbed water formed on the mineral surface, and the film thickness increased with time as the forsterite began to dissolve. However, after approximately 2.5 hours, the trend reversed, and a carbonate precipitate began to form on the forsterite surface, exposing dramatic chemical changes in the thin-water film. Collectively, these applications illustrate how the high-pressure IR titration system can provide molecular-level information about the interactions between variably wet scCO2 and minerals relevant to underground storage of CO2 (geologic carbon sequestration). The apparatus could also be utilized to study high-pressure interfacial chemistry in other areas such as catalysis, polymerization, food processing, and oil and gas recovery.

Research Organization:
Pacific Northwest National Lab. (PNNL), Richland, WA (United States). Environmental Molecular Sciences Lab. (EMSL)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC05-76RL01830
OSTI ID:
1159012
Report Number(s):
PNNL-SA-101170; 47427; 47824; AA7020000
Journal Information:
Review of Scientific Instruments, 85(4):Article No. 044102, Journal Name: Review of Scientific Instruments, 85(4):Article No. 044102