skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Precision timing measurement of phototube pulses using a flash analog-to-digital converter

Journal Article · · Nuclear Instruments and Methods A

We present the timing characteristics of the flash ADC readout of the GlueX forward calorimeter, which depends on precise measurement of arrival time of pulses from FEU 84-3 photomultiplier tubes to suppress backgrounds. The tests presented were performed using two different 250 MHz prototype flash ADC devices, one with eight-bit and one with 12-bit sampling depth. All measured time resolutions were better than 1 ns, independent of signal size, which is the design goal for the GlueX forward calorimeter. For pulses with an amplitude of 100 mV the timing resolution is 0.57±0.18 ns, while for 500 mV pulses it is 0.24±0.08 ns.

Research Organization:
Thomas Jefferson National Accelerator Facility (TJNAF), Newport News, VA (United States)
Sponsoring Organization:
USDOE Office of Science (SC)
DOE Contract Number:
AC05-06OR23177
OSTI ID:
1008175
Report Number(s):
JLAB-PHY-10-1315; DOE/OR/23177-1489; TRN: US201106%%318
Journal Information:
Nuclear Instruments and Methods A, Vol. 622, Issue 1; ISSN 0168-9002
Country of Publication:
United States
Language:
English