Available on CMS information server CMS IN 2005/006 CMS Internal Note
 

Summary: Available on CMS information server CMS IN 2005/006
CMS Internal Note
The content of this note is intended for CMS internal use and distribution only
January 27, 2005
Variation of relative gains as a function of HV for
the Hamamatsu R7525HA PMTs used in the HF
Forward Calorimeter
E. GĻulmez
Bogazici University, Bebek 34342 Istanbul, Turkey
U. Akgun and Y.Onel
University of Iowa, Iowa City 52242 IA, USA
Abstract
Relative gains of the Hamamatsu R7525HA PMTs used in the HF Forward Calorimeter were studied
at different PMT high voltages on a sample of 223 PMTs. Relative gains do not vary much but the
overall uncertainty increases slightly. There is an additional 5% uncertainty at 1350 V (HV used
during the test beam 2004).
1 Introduction
Relative gains at 1100 V of all the PMTs to be used in the HF Forward Calorimeter were measured at the University
of Iowa PMT test sation. However, relative gains at different HV settings may not necessarily be the same as that
at 1100 V because of the different gain characteristics of these PMTs (absolute gains vary by a factor of five). We

  