skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: An automated instrument for controlled-potential coulometry: System documentation

Technical Report ·
DOI:https://doi.org/10.2172/6772203· OSTI ID:6772203

An automated controlled-potential coulometer has been developed at the Savannah River Plant for the determination of plutonium. Two such coulometers have been assembled, evaluated, and applied. The software is based upon the methodology used at the Savannah River Plant, however the system is applicable with minimal software modifications to any of the methodologies used throughout the nuclear industry. These state-of-the-art coulometers feature electrical calibration of the integration system, background current corrections, and control-potential adjustment capabilities. Measurement precision within 0.1% has been demonstrated. The systems have also been successfully applied to the determination of pure neptunium solutions. The design and documentation of the automated instrument are described herein. Each individual module's operation, wiring layout, and alignment are described. Interconnection of the modules and system calibration are discussed. A complete set of system prints and a list of associated parts are included. 9 refs., 10 figs., 6 tabs.

Research Organization:
Savannah River Site (SRS), Aiken, SC (United States). Savannah River National Lab. (SRNL)
DOE Contract Number:
AC09-76SR00001
OSTI ID:
6772203
Report Number(s):
DP-1751; ON: DE89001029
Resource Relation:
Other Information: Paper copy only, copy does not permit microfiche production
Country of Publication:
United States
Language:
English