We have measured the responses to /sup 137/Cs (662 keV) of both a 1-inch-diam by 2-inch-thick NaI(Tl) scintillator optically coupled to a 1-inch-diam by 800-..mu..m-thick mercuric iodide (HgI/sub 2/) photodetector, and a 1-cm-diam by 1-cm-thick CaWO/sub 4/ scintillator coupled to a 1.3-cm-diam by 600-..mu..m-thick HgI/sub 2/ photodetector. Best spectral resolution to /sup 137/Cs was 7.8% FWHM for the NaI(Tl)-HgI/sub 2/ and 12.5% FWHM for the CaWO/sub 4/-HgI/sub 2/ detectors; peak-to-valley ratios were 26:1 and 16:1, respectively. HgI/sub 2/ detectors operate at room temperature and their use in scintillation spectroscopy presents the ultimate miniaturization of scintillation detectors, limited mainly by the size of the scintillation crystal.

Research Organization:
EG and G, Inc., Goleta, CA (USA). Energy Measurements Group; University of Southern California, Marina Del Rey (USA). Inst. for Physics and Imaging Science
DOE Contract Number:
AC08-83NV10282
OSTI ID:
6416379
Report Number(s):
DOE/NV/10282-2038; CONF-841007-6; ON: DE85001437
Resource Relation:
Conference: Nuclear science symposium, Orlando, FL, USA, 31 Oct 1984; Other Information: Portions are illegible in microfiche products
Country of Publication:
United States
Language:
English