Title: Background radiation accumulation and lower limit of detection in thermoluminescent beta-gamma dosimeters used by the centralized external dosimetry system

Technical Report ·
DOI:https://doi.org/10.2172/5948905· OSTI ID:5948905

A value for average background radiation'' of 0.75 mR/week has been determined from a total of 1680 thermoluminescent dosimeters (TLD's) exposed in 70 houses for periods up to one year. The distribution of results indicates a rather large variation among houses, with a few locations exhibiting backgrounds double the general average. Some discrepancies in the short-term background accumulation of TLD's have been explained as being due to light leakage through the dosimeter cases. In addition the lower limit of detection (L{sub D}) for deep and shallow dose equivalents has been determined for these dosimeters. The L{sub D} for occupational exposure depends strongly on the time a dosimeter is exposed to background radiation in the field. The L{sub D} can vary from a low of 2.4 mrem for high energy gamma rays when the background accumulation period is less than a few weeks to values as high as 66 mrem for uranium beta particles when background has been allowed to accumulate for more than 21 weeks.

Research Organization:
Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States)
Sponsoring Organization:
USDOE; USDOE, Washington, DC (United States)
DOE Contract Number:
AC05-84OR21400
OSTI ID:
5948905
Report Number(s):
ORNL/TM-11995; ON: DE92005027