Title: Charge division using carbon filaments for obtaining coordinate information from detection of single electrons

Conference · · IEEE Trans. Nucl. Sci.; (United States)
OSTI ID:5823660

Seven micron diameter carbon filaments forming the anode of a multiwire proportional chamber have been used to detect single electrons. Charge division techniques applied to the 5 cm long wire resulted in a position resolution of sigma/L < 2% for a collected signal charge of 30 fC.

Research Organization:
Stanford Linear Accelerator Center, Stanford University, Stanford, CA 94305
DOE Contract Number:
AC03-76SF00515
OSTI ID:
5823660
Report Number(s):
CONF-851009-
Journal Information:
IEEE Trans. Nucl. Sci.; (United States), Vol. NS-33:1; Conference: IEEE nuclear science symposium, San Francisco, CA, USA, 23 Oct 1985
Country of Publication:
United States