Title: Nonlinear partial volume artifacts in x-ray computed tomography

Journal Article · · Med. Phys.; (United States)
DOI:https://doi.org/10.1118/1.594678· OSTI ID:5579913

Dark streaks connecting the petrous bones are often observed in cranial transverse section CT reconstructions. These artifacts are usually only slightly diminished by two-pass beam hardening corrections. However, it is found that by narrowing the slice thickness the artifacts are substantially reduced. In this paper, it is shown that axial partial volume effects can account for the presence of the artifacts. These axial partial volume effects occur when, at any point in the slice, the object has axial variations in attenuation. In such cases the logarithm of the integrated intensity measured by the detector is not a linear function of the integrated attenuation (even for monochromatic beams). This nonlinearity causes inconsistencies in the data set which in turn can cause streaks in the image. We have studied the partial volume effect using computer simulation. Algorithms are presented whose purpose is to correct for these effects by estimating the axial variation using neighboring slices. These correction algorithms are successful in computer simulation cases but failed with clinical data. It is concluded that no practical correction method is viable unless overlap scanning is employed. However, thin-slice scanning for sections where these artifacts are common is perhaps a more preferable solution.

Research Organization:
General Electric Co., Milwaukee, WI
OSTI ID:
5579913
Journal Information:
Med. Phys.; (United States), Vol. 7:3
Country of Publication:
United States
Language:
English