skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: SOURCE OF ERROR IN NEUTRON PERSONNEL DOSIMETRY WITH NUCLEAR TRACK EMULSION (in German)

Journal Article · · Atomkernenergie (West Germany) Merged with Kerntechnik to form Atomkernenerg./Kerntech. Acta Radiol. Changed to Acta Radiol.: Oncol., Radiat. Phys.

The accuracy of the neutron rem dose determination by nuclear track emulsion techniques in personnel dosimetry in the case of the most used Kodak NTA, Type A, emulsion between 0.02 and 20 rem and 0.6 and 20 Mev hardly depends on the directional dependence of the film, subjective counting errors besides, if the gamma dose does not exceed 5 to 10 r, neither on the fog granularity. The influence of less exact calibration and energy dependence of the neutron sensitivity as well as the orientation of the dosimeter to the neutron source and the human body and the fading effects are more serious. The fading is studied as a function of relative humidity at 20 deg C and it is found, that in high relative humidity after several days, nuclear tracks are no more countable. Suitable packing of the film can reduce this effect. One should take care of the evaluation of neutron dose measurements. (auth)

Research Organization:
Kernforschungsanlage, Julich, Ger.
NSA Number:
NSA-17-016434
OSTI ID:
4754077
Journal Information:
Atomkernenergie (West Germany) Merged with Kerntechnik to form Atomkernenerg./Kerntech. Acta Radiol. Changed to Acta Radiol.: Oncol., Radiat. Phys., Vol. Vol: 8; Other Information: Orig. Receipt Date: 31-DEC-63
Country of Publication:
Country unknown/Code not available
Language:
German