Preamplifier and AD converter system for infrared focal plane arrays
We describe a precision preamplifier and analog-to-digital (AD) converter system for use with infrared focal plane arrays. The system uses a correlated double sampling integrated circuit which performs an analog subtraction of each full pixel level from its respective reset level prior to digitization by a 14-bit AD converter; this process is also known as ''delta-reset'' sampling. The system was designed to have good immunity to electrical noise interference and it has a measured dynamic range of better than 16384. A four channel version of this system is used to read out a 128x128 pixel Si:As array at a rate of 2.4 {mu}s/pixel for the Long Wavelength InfraRed Camera a 7--13 {mu}m camera designed and built for the Keck 10 meter telescope.
- Sponsoring Organization:
- (US)
- OSTI ID:
- 40204895
- Journal Information:
- Review of Scientific Instruments, Vol. 71, Issue 10; Other Information: DOI: 10.1063/1.1311943; Othernumber: RSINAK000071000010003921000001; 007011RSI; PBD: Oct 2000; ISSN 0034-6748
- Publisher:
- The American Physical Society
- Country of Publication:
- United States
- Language:
- English
Similar Records
Long wavelength infrared camera (LWIRC): a 10 micron camera for the Keck Telescope
Megapixel imaging camera for expanded H{sup {minus}} beam measurements