Title: An imaging diode array soft x-ray diagnostic for Z (abstract)

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.1326022· OSTI ID:40204803

Measurements of the hohlraum wall temperature in Z-pinch driven hohlraum experiments require looking through small (2--4 mm diameter) diagnostic holes that undergo some degree of hole closure. The existing soft x-ray diagnostics on Z measure the total flux exiting this diagnostic hole and are therefore affected by this hole closure. To avoid having to measure the effective diagnostic hole area we have designed and constructed an imaging diode array (IDA) that incorporates pinhole imaging and an array of filtered silicon diodes to measure the absolute x-ray intensity from a spatially resolved region of a target. The instrument uses silicon diodes with subnanosecond time response that are sensitive to soft x rays in the range 100--3000 eV. An image of the target area is projected onto the silicon diodes using pinholes. Between each pinhole and its respective diode is a soft x-ray filter. The material and thickness of the filter are selected to allow unfolding of spectral information in the 100--3000 eV spectral region. We plan to insert a set of grazing-incidence mirrors between each of the filter/diode pairs in a future version of this instrument to better define the spectral bandpass of each diode channel. Radiation from the target region is monitored by a gated microchannel-plate-intensified image recording device that is located immediately behind the diode array. A small shadow in the recorded image corresponds to the specific area of the target that is imaged onto each silicon diode. We are presently fielding this instrument in experiments on the Z facility located at Sandia National Laboratories in Albuquerque, NM. The instrument is located on the same line-of-sight and measures the same spatial region as a filtered fast-framing x-ray pinhole camera and a transmission grating spectrometer. This article describes the design of the IDA diagnostic and presents the results of measurements obtained in hohlraum experiments conducted on Z.

Sponsoring Organization:
(US)
OSTI ID:
40204803
Journal Information:
Review of Scientific Instruments, Vol. 72, Issue 1; Other Information: Othernumber: RSINAK000072000001001223000001; 748101CON; PBD: Jan 2001; ISSN 0034-6748
Publisher:
The American Physical Society
Country of Publication:
United States
Language:
English

Similar Records

Characterization of axially directed x rays generated from a target within a high-power z-pinch (abstract)
Journal Article · Mon Jan 01 00:00:00 EST 2001 · Review of Scientific Instruments · OSTI ID:40204803

Measurement of the energy and power radiated by a pulsed blackbody x-ray source.
Journal Article · Wed Feb 01 00:00:00 EST 2006 · Proposed for publication in Physical Review E. · OSTI ID:40204803

Diagnostics on Z (invited)
Journal Article · Mon Jan 01 00:00:00 EST 2001 · Review of Scientific Instruments · OSTI ID:40204803