SU-F-T-66: Characteristics of Electron Beams From Varian Trubeam
- University Pennsylvania, Philadelphia, PA (United States)
Purpose: The purpose of this study was to compare the electron beam data between Truebeam and 2300ix Varian accelerators for percent depth dose for broad beam and small circular cutouts, cone factors, head scatter factor as a function of cone size and SSD, phantom scatter factor, blocking factor, distance factor and virtual source position. Methods: Measurements were performed for Truebeam and 2300ix Varian accelerators. The main energies used were: 6, 9, 12, 16 and 20 MeV. PDD was measured at SSD = 100 cm for open beam and small circular cutouts (r = 0.5, 1.0, 1.5, 2.0, 3.0, 4.0 and 6.6cm) for different energies. Measurements to determine the head scatter factor (H) were done as a function of radius for six representative energies and five cone sizes (6, 10, 15, 20 and 25cm2). The phantom scatter factor (PSF) is defined as the ratio of blocking factor in water at reference depth and head scatter factor in air. PSF was measured as a function of radius and electron energy. Distance factor was measured for all energies and cones for three SSD’s (100, 110 and 120cm). Results: The percent depth dose (PDD) was measured for small cutouts of radius r = 0.5, 1.0, 1.5, 2.0, 3.0, 4.0, 5.0 and 5.6cm. Blocking factor (BF) was measured for Truebeam and 2300ix accelerators, for different circular cutouts and energies for a 10×10 cone. Cone factors were compared between the two accelerators for different energies and applicator sizes. Conclusion: Cone factors measured for the two accelerator types differ by up to 5% for the largest applicator size. Blocking factors differs by up to 3%, with the largest variation for the smallest field size (0.5cm). Distance factor for different SSD’s differ by up to 4.5%.
- OSTI ID:
- 22642314
- Journal Information:
- Medical Physics, Vol. 43, Issue 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
- Country of Publication:
- United States
- Language:
- English
Similar Records
Output calculation of electron therapy at extended SSD using an improved LBR method
SU-E-T-601: Output Factor Calculation of Irregular Shape Electron Cutout at Extended SSD