skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: SU-E-T-81: A Study On Correlation Between Gamma Analysis for Midline and Lateralized Tumors Using VMAT

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4924442· OSTI ID:22545210
 [1];  [2]
  1. Malabar Cancer Centre, Kannur, Kerala (India)
  2. University of Calicut, Calicut, Kerala (India)

Purpose: To evaluate the fluence for the midline and lateralized tumors for VMAT technique using 2D seven29 detector array combined with the Octavius phantom. Methods: 60 cases that are already being treated with volumetric modulated arc therapy (VMAT) have selected for this study. This includes tumors situated at the medial and lateral. Medial refers to the tumor situated at the midline of the body and lateral means toward the side or away from the midline of the body. Verification plans were created for each treatment plan in Varian Eclipse treatment planning system (version10, Varian medical systems, Palo Alto,CA) with the 2D Seven29 detector array and the Octavius phantom(PTW, Freiburg, Germany). Measurements were performed on a Varian Clinac 2100 iX, linear accelerator equipped with a millennium 120 leaf collimator. Analysis was done by comparing the fluence measured for the tumors situated on the midline and tumors situated laterally. Results: Fluence measured for all the delivered plans were analyzed using Verisoft software (PTW, Freiburg, Germany). The gamma pass percentage for midline tumors were found to be higher compared with the lateralized ones. The standard deviation between gamma values for midline and lateralized tumors is 2.18 and 3.5 respectively. Also the standard deviation between the point doses for midline and lateralized tumors is 0.38 and 0.29 respectively. The average gamma passing rate for midline tumors is 96.55% and for lateralized tumors are 94.94% for 3%DD and 3mm DTA criteria. From the T test, it was found that there is no significant difference between the gamma pass percentage between midline and lateralized tumors with p value of 0.28. Conclusion: There is no particular correlation found in the gamma pass criteria for midline and lateralized tumors.

OSTI ID:
22545210
Journal Information:
Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English