skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: SU-E-T-589: Optimization of Patient Head Angle Position to Spare Hippocampus During the Brain Radiation Therapy

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4924952· OSTI ID:22496302
;  [1]; ; ; ;  [2]
  1. Radiation Oncology, Seoul St. Mary’s Hospital, Seoul (Korea, Republic of)
  2. The Catholic University of Korea, Seoul (Korea, Republic of)

Purpose: Hippocampus is one of the important organs which controls emotions, behaviors, movements the memorizing and learning ability. In the conventional head & neck therapy position, it is difficult to perform the hippocampal-sparing brain radiation therapy. The purpose of this study is to investigate optimal head angle which can save the hippocampal-sparing and organ at risk (OAR) in conformal radiation therapy (CRT), Intensity modulation radiation therapy (IMRT) and helical tomotherapy (HT). Methods: Three types of radiation treatment plans, CRT, IMRT and Tomotherapy plans, were performed for 10 brain tumor patients. The image fusion between CT and MRI data were used in the contour due to the limited delineation of the target and OAR in the CT scan. The optimal condition plan was determined by comparing the dosimetric performance of the each plan with the use of various parameters which include three different techniques (CRT, IMRT, HT) and 4 angle (0, 15, 30, 40 degree). The each treatment plans of three different techniques were compared with the following parameters: conformity index (CI), homogeneity index (HI), target coverage, dose in the OARs, monitor units (MU), beam on time and the normal tissue complication probability (NTCP). Results: HI, CI and target coverage was most excellent in head angle 30 degree among all angle. When compared by modality, target coverage and CI showed good results in IMRT and TOMO than compared to the CRT. HI at the head angle 0 degrees is 1.137±0.17 (CRT), 1.085±0.09 (IMRT) and 1.077±0.06 (HT). HI at the head angle 30 degrees is 1.056±0.08 (CRT), 1.020±0.05 (IMRT) and 1.022±0.07 (HT). Conclusion: The results of our study show that when head angle tilted at 30 degree, target coverage, HI, CI were improved, and the dose delivered to OAR was reduced compared with conventional supine position in brain radiation therapy. This work was supported by the Radiation Technology R&D program (No. 2013M2A2A7043498) and the Mid-career Researcher Program (2014R1A2A1A10050270) through the National Research Foundation of Korea funded by the Ministry of Science, ICT&Future Planning.

OSTI ID:
22496302
Journal Information:
Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English