Novel 4x4 SiPM Array Readout with Integrated Preamplification Stage, Optimized for the PWO Detectors of the EIC EEEMCal [Slides]
- Crytur USA, Watertown, MA (United States)
- The Catholic Univ. of America, Washington, DC (United States)
- Thomas Jefferson National Accelerator Facility (TJNAF), Newport News, VA (United States)
We are reporting on a new readout circuit developed for the lead tungstate (PbWO4) scintillation detectors for the Electron Ion Collider (EIC) Electron Endcap Electromagnetic Calorimeter EEEMCal. The high magnetic field region precludes the use of photomultiplier tubes while the detector requirements specify good spectral resolution performance and fast pulse integration over a large dynamic range. We selected a silicon photomultiplier (SiPM) from Hamamatsu and produced a matrix of 4 x 4 sensors to cover the 20 mm x 20 mm scintillator coupling surface. Signal acquisition and amplification electronics boards were designed and integrated with the sensor board to produce a compact high performance readout package. A prototype detector was built and tested at Jefferson Lab with encouraging resolution and timing performance results.
- Research Organization:
- Thomas Jefferson National Accelerator Facility (TJNAF), Newport News, VA (United States)
- Sponsoring Organization:
- USDOE Office of Science (SC), Nuclear Physics (NP)
- DOE Contract Number:
- AC05-06OR23177
- OSTI ID:
- 2246761
- Report Number(s):
- JLAB-PHY-23-3980; DOE/OR/23177-7317
- Resource Relation:
- Conference: 19. 2023 Symposium on Radiation Measurements and Applications (SORMA XIX), Ann Arbor, MI (United States), 22 May 2023; Related Information: https://rma-symposium.engin.umich.edu/home
- Country of Publication:
- United States
- Language:
- English