Title: How MRI Compatible is 'MRI Compatible'? A Systematic Comparison of Artifacts Caused by Biopsy Needles at 3.0 and 1.5 T

Journal Article · · Cardiovascular and Interventional Radiology
 [1];  [2];  [1];  [3];  [1]
  1. RWTH Aachen University Hospital, Department of Diagnostic and Interventional Radiology (Germany)
  2. Klinikum Osnabrueck, Roentgen- und Strahlenklinik (Germany)
  3. Justus-Liebig University Giessen, Department of Radiology (Germany)

Purpose: This study was designed to systematically investigate artifacts caused by interventional needles recommended for use in MRI, with focus on field strength, needle/mandrin type, orientation and sequence. Methods: Eight different MRI compatible needles were placed in porcine tissue and examined at 1.5 and 3.0 T with balanced-steady-state-free-precession (B-SSFP) and T1-weighted-spoiled-gradient-echo (T1-SPGR) sequences in different orientations to B{sub 0}. Artifact diameters with regards to the primary, inner, and secondary, outer artifacts were assessed and statistically evaluated. Results: The types and degree of artifacts varied considerably, especially between different mandrin types even for the same needles. Orientation of the needle in the magnetic field was another main contributor to the artifact dimensions. Less important factors were the type of pulse sequence and field strength. Artifacts ranged from 0.7 mm (steel, 0 Degree-Sign , B-SSFP, 3.0 T, inner) to 71.4 mm (nitinol, 90 Degree-Sign , B-SSFP, 1.5 T, outer). Inner artifact diameters in B-SSFP were slightly larger (8.2 {+-} 5.7 mm) than those in T1-SPGR (7.6 {+-} 5.4 mm) and comparable between 1.5 and 3.0 T (e.g., 8.0 vs. 8.4 mm, B-SSFP). Conclusions: Although all were sold as 'MR compatible,' the artifacts differed greatly between needle types, and even more so for different mandrins. The results suggest an empirical approach to the needle choice based on lesion type and approach angle.

OSTI ID:
22207965
Journal Information:
Cardiovascular and Interventional Radiology, Vol. 36, Issue 6; Other Information: Copyright (c) 2013 Springer Science+Business Media New York and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE); http://www.springer-ny.com; Country of input: International Atomic Energy Agency (IAEA); ISSN 0174-1551
Country of Publication:
United States
Language:
English