Title: The inter- and intrafraction reproducibilities of three common IMRT delivery techniques

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.3476413· OSTI ID:22098530
; ;  [1]
  1. Department of Radiology and Department of Radiation Oncology, University of Texas Health Science Center at San Antonio, San Antonio, Texas 78229 (United States)

Purpose: Intensity modulated radiation therapy (IMRT) treatment delivery requires higher precision than conventional 3D treatment delivery because of the sensitivity of the resulting dose to small geometric misalignment of the modulated beamlets. The chosen treatment delivery technique will affect the treatment precision in different ways, based on the characteristics of the delivery method. Delivery using a multileaf collimator (MLC) can reduce treatment time and therapist workload, but typically requires a greater number of monitor units and the fields are prone to both systematic and random leaf positioning errors. An alternative to MLC-based fields, patient specific brass compensators, do not suffer from these leaf positioning errors. In our study, we set out to investigate which delivery method will provide the highest levels of dosimetric reproducibility and the minimum amount of interfraction variability. Methods: In our study, a seven field IMRT plan for a head and neck treatment was created using the Pinnacle{sup 3} treatment planning system and the intensity maps for each field were obtained. The intensity maps of the fields were delivered with a Varian 2100C/D linear accelerator, using solid compensators and sliding window (SW) and step-and-shoot (SS) MLC segments. Three fields were selected from the seven-beam IMRT plan for comparison. Analysis was carried out using the MatriXX ion chamber array, radiochromic film, and Varian dynalog files. Results: Our results show that the error in MLC leaf positioning has no gantry angle dependence. The compensator and SW deliveries showed excellent agreement, even when stricter than usual gamma criteria were applied. However, we noted that under these strict conditions, the SS fields had at least ten times more pixels out of range than did the compensators. When using step-and-shoot MLC fields, it was observed that the increase in dose rate or the increase of MU/segment degrades the quality of the plan. Analysis of the dynalog files showed that while each individual field had its own propensity for error, all fields showed the same trend: a greater percentage of time the leaves are out of position as dose rate increases, MUs decrease, or both. Conclusions: The compensator-based field and both types of MLC-based fields have MatriXX results that are within the clinically acceptable tolerance of 3% dose difference and 2 mm DTA. However, when the criteria are tightened, it becomes evident that the compensators have a definite advantage over their comparable MLC-based competitors in terms of interfraction reproducibility. Fewer monitor units are required to deliver each portal, potentially improving patient outcomes and reducing unwanted side effects to both patients and therapists. In centers without MLC, compensators represent a simple and cost effective way to offer patients state of the art treatment. Based on the results of this study, compensator-based IMRT is a reliable, viable option for use in clinics both with and without MLC-equipped linacs.

OSTI ID:
22098530
Journal Information:
Medical Physics, Vol. 37, Issue 9; Other Information: (c) 2010 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English