Title: Photoconductive detectors with fast temporal response for laser produced plasma experiments

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.2955614· OSTI ID:21266469
; ; ; ;  [1];  [2]
  1. Lawrence Livermore National Laboratory, 7000 East Ave., Livermore, California 94550 (United States)
  2. National Security Technologies LLC, P.O. Box 2710, Livermore, California 94550 (United States)

Processes during laser plasma experiments typically have time scales that are less than 100 ps. The measurement of these processes requires x-ray detectors with fast temporal resolution. We have measured the temporal responses and linearity of several different x-ray sensitive photoconductive detectors (PCDs). The active elements of the detectors investigated include both diamond (natural and synthetic) and GaAs crystals. The typical time responses of the GaAs PCDs are approximately 60 ps, respectively. Some characterizations using x-ray radiation from a synchrotron radiation source are presented.

OSTI ID:
21266469
Journal Information:
Review of Scientific Instruments, Vol. 79, Issue 10; Other Information: DOI: 10.1063/1.2955614; (c) 2008 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English