Aqueous blackbody calibration source for millimeter-wave/terahertz metrology
This paper describes a calibrated broadband emitter for the millimeter-wave through terahertz frequency regime, called the aqueous blackbody calibration source. Due to its extremely high absorption, liquid water is chosen as the emitter on the basis of reciprocity. The water is constrained to a specific shape (an optical trap geometry) in an expanded polystyrene (EPS) container and maintained at a selected, uniform temperature. Uncertainty in the selected radiometric temperature due to the undesirable reflectance present at a water interface is minimized by the trap geometry, ensuring that radiation incident on the entrance aperture encounters a pair of s and a pair of p reflections at 45 deg. . For water reflectance Rw of 40% at 45 deg. in W-band, this implies a theoretical effective aperture emissivity of (1-R{sup 2}wsR{sup 2}wp)>98.8%. From W-band to 450 GHz, the maximum radiometric temperature uncertainty is {+-}0.40 K, independent of water temperature. Uncertainty from 450 GHz to 1 THz is increased due to EPS scattering and absorption, resulting in a maximum uncertainty of -3 K at 1 THz.
- OSTI ID:
- 21175679
- Journal Information:
- Applied Optics, Vol. 47, Issue 30; Other Information: DOI: 10.1364/AO.47.005604; (c) 2008 Optical Society of America; Country of input: International Atomic Energy Agency (IAEA); ISSN 0003-6935
- Country of Publication:
- United States
- Language:
- English
Similar Records
WVR-GPS comparison measurements and calibration of the 20-32 GHz tropospheric water vapor absorption model.
A liquid-helium-cooled absolute reference cold load for long-wavelength radiometric calibration