Title: A Dosimetric Comparison of Accelerated Partial Breast Irradiation Techniques: Multicatheter Interstitial Brachytherapy, Three-Dimensional Conformal Radiotherapy, and Supine Versus Prone Helical Tomotherapy

Journal Article · · International Journal of Radiation Oncology, Biology and Physics
 [1];  [2];  [1];  [1]
  1. Department of Human Oncology, University of Wisconsin, Madison, WI (United States)
  2. Department of Medical Physics, University of Wisconsin, Madison, WI (United States)

Purpose: To compare dosimetrically four different techniques of accelerated partial breast irradiation (APBI) in the same patient. Methods and Materials: Thirteen post-lumpectomy interstitial brachytherapy (IB) patients underwent imaging with preimplant computed tomography (CT) in the prone and supine position. These CT scans were then used to generate three-dimensional conformal radiotherapy (3D-CRT) and prone and supine helical tomotherapy (PT and ST, respectively) APBI plans and compared with the treated IB plans. Dose-volume histogram analysis and the mean dose (NTD{sub mean}) values were compared. Results: Planning target volume coverage was excellent for all methods. Statistical significance was considered to be a p value <0.05. The mean V100 was significantly lower for IB (12% vs. 15% for PT, 18% for ST, and 26% for 3D-CRT). A greater significant differential was seen when comparing V50 with mean values of 24%, 43%, 47%, and 52% for IB, PT, ST, and 3D-CRT, respectively. The IB and PT were similar and delivered an average lung NTD{sub mean} dose of 1.3 Gy{sub 3} and 1.2 Gy{sub 3}, respectively. Both of these methods were statistically significantly lower than the supine external beam techniques. Overall, all four methods yielded similar low doses to the heart. Conclusions: The use of IB and PT resulted in greater normal tissue sparing (especially ipsilateral breast and lung) than the use of supine external beam techniques of 3D-CRT or ST. However, the choice of APBI technique must be tailored to the patient's anatomy, lumpectomy cavity location, and overall treatment goals.

OSTI ID:
20951725
Journal Information:
International Journal of Radiation Oncology, Biology and Physics, Vol. 68, Issue 3; Other Information: DOI: 10.1016/j.ijrobp.2007.03.005; PII: S0360-3016(07)00391-4; Copyright (c) 2007 Elsevier Science B.V., Amsterdam, Netherlands, All rights reserved; Country of input: International Atomic Energy Agency (IAEA); ISSN 0360-3016
Country of Publication:
United States
Language:
English