Title: Improved outcome with novel device for low-pressure PTCA in de novo and in-stent lesions

Journal Article · · Cardiovascular Radiation Medicine

Purpose: Complex lesion morphology requiring the use of high pressure to effect lumen expansion and in-stent restenosis (ISR) remain two indications that challenge conventional PTCA balloons. We report on a new PTCA device that is designed to provide precise, low-pressure dilatation of both de novo and in-stent lesions. Methods: The FX miniRAIL catheter (FX) has an integral wire positioned external to a dilating balloon and a short, 12-mm guidewire lumen distal to the balloon. The balloon inflates against the guidewire and the external wire to prevent slippage and to introduce high focal longitudinal stresses at low inflation pressures. In this initial study, the FX was used in 37 lesions (25 de novo, 12 in-stent; vessel reference diameter=2.73{+-}0.49 mm) in 30 patients. A stepwise inflation protocol and QCA were used to determine the balloon pressure at which the stenosis was resolved (stenosis resolution pressure, SRP). Results: All lesions (100%) were easily reached, crossed and dilated without complication. The SRP was 4.5{+-}2.9 atm, and no balloon slippage was observed. Residual stenosis after FX was 26.39{+-}13.29%. Minor dissections (Types A and B) were observed in eight lesions (21.6%). Target lesion revascularization (TLR) and target vessel revascularization (TVR) at follow-up (8.1{+-}1.5 months) were 8.3% and 12.5%, respectively. Conclusion: The design of the FX is versatile and appears to provide for a safe, effective and improved low-pressure PTCA technique in de novo and in-stent lesions.

OSTI ID:
20650551
Journal Information:
Cardiovascular Radiation Medicine, Vol. 4, Issue 1; Other Information: DOI: 10.1016/S1522-1865(03)00118-5; PII: S1522186503001185; Copyright (c) 2003 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved; Country of input: International Atomic Energy Agency (IAEA); ISSN 1522-1865
Country of Publication:
United States
Language:
English

Similar Records

Beta-radiation therapy for long lesions in native coronary vessels A matched comparison between de novo and in-stent restenotic lesions
Journal Article · Sat Mar 01 00:00:00 EST 2003 · Cardiovascular Radiation Medicine · OSTI ID:20650551

Safety and efficacy of manual stepping and overlapping of {beta}-emitter for diffuse in-stent restenosis lesions
Journal Article · Fri Mar 01 00:00:00 EST 2002 · Cardiovascular Radiation Medicine · OSTI ID:20650551

One Year Clinical Outcomes of Renal Artery Stenting: The Results of ODORI Registry
Journal Article · Tue Jun 15 00:00:00 EDT 2010 · Cardiovascular and Interventional Radiology · OSTI ID:20650551