High resolution transmission grating spectrometer for edge toroidal rotation measurements of tokamak plasmas
- University of California, Davis, California 95616 (United States)
We present a high throughput (f/3) visible (3500-7000 A) Doppler spectrometer for toroidal rotation velocity measurements of the Alcator C-Mod tokamak plasma. The spectrometer has a temporal response of 1 ms and a rotation velocity sensitivity of {approx}10{sup 5} cm/s. This diagnostic will have a tangential view and map out the plasma rotation at several locations along the outer half of the minor radius (r/a>0.5). The plasma rotation will be determined from the Doppler shifted wavelengths of D{sub {alpha}} and magnetic and electric dipole transitions of highly ionized impurities in the plasma. The fast time resolution and high spectral resolving power are possible due to a 6 in. diam circular transmission grating that is capable of {lambda}/{delta}{lambda}{approx}15 500 at 5769 A in conjunction with a 50 {mu}m slit.
- OSTI ID:
- 20641262
- Journal Information:
- Review of Scientific Instruments, Vol. 75, Issue 10; Conference: 15. topical conference on high temperature plasma diagnostics, San Diego, CA (United States), 19-22 Apr 2004; Other Information: DOI: 10.1063/1.1789591; (c) 2004 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748