Title: A method to test the reproducibility and to improve performance of computer-aided detection schemes for digitized mammograms

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.1806291· OSTI ID:20634417
; ; ;  [1]
  1. Department of Radiology, University of Pittsburgh, 300 Halket St, Suite 4200, Pittsburgh, Pennsylvania 15213-3180 (United States)

The purpose of this study is to develop a new method for assessment of the reproducibility of computer-aided detection (CAD) schemes for digitized mammograms and to evaluate the possibility of using the implemented approach for improving CAD performance. Two thousand digitized mammograms (representing 500 cases) with 300 depicted verified masses were selected in the study. Series of images were generated for each digitized image by resampling after a series of slight image rotations. A CAD scheme developed in our laboratory was applied to all images to detect suspicious mass regions. We evaluated the reproducibility of the scheme using the detection sensitivity and false-positive rates for the original and resampled images. We also explored the possibility of improving CAD performance using three methods of combining results from the original and resampled images, including simple grouping, averaging output scores, and averaging output scores after grouping. The CAD scheme generated a detection score (from 0 to 1) for each identified suspicious region. A region with a detection score >0.5 was considered as positive. The CAD scheme detected 238 masses (79.3% case-based sensitivity) and identified 1093 false-positive regions (average 0.55 per image) in the original image dataset. In eleven repeated tests using original and ten sets of rotated and resampled images, the scheme detected a maximum of 271 masses and identified as many as 2359 false-positive regions. Two hundred and eighteen masses (80.4%) and 618 false-positive regions (26.2%) were detected in all 11 sets of images. Combining detection results improved reproducibility and the overall CAD performance. In the range of an average false-positive detection rate between 0.5 and 1 per image, the sensitivity of the scheme could be increased approximately 5% after averaging the scores of the regions detected in at least four images. At low false-positive rate (e.g., {<=}average 0.3 per image), the grouping method alone could increase CAD sensitivity by 7%. The study demonstrated that reproducibility of a CAD scheme can be tested using a set of slightly rotated and resampled images. Because the reproducibility of true-positive detections is generally higher than that of false-positive detections, combining detection results generated from subsets of rotated and resampled images could improve both reproducibility and overall performance of CAD schemes.

OSTI ID:
20634417
Journal Information:
Medical Physics, Vol. 31, Issue 11; Other Information: DOI: 10.1118/1.1806291; (c) 2004 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English