Title: Thermoacoustic range verification in the presence of acoustic heterogeneity and soundspeed errors - Robustness relative to ultrasound image of underlying anatomy

Journal Article · · Medical Physics
DOI:https://doi.org/10.1002/mp.13256· OSTI ID:1503258
 [1];  [2];  [2]
  1. Univ. of Wisconsin, Milwaukee, WI (United States). Dept. of Physics
  2. Argonne National Lab. (ANL), Lemont, IL (United States)

Purpose: To demonstrate robustness of thermooacoustic range verification to acoustic heterogeneity and discrepancies between assumed and true propagation speed, i.e., soundspeed errors. Methods: A beam sweeper was used to deliver 250 ns pulses that deposited 0.26 Gy of 16 MeV protons and 2.3 Gy of 60 MeV helium ions into water and oil targets, respectively. Thermoacoustic signals were detected by a 96-channel ultrasound array with a 1-4 MHz sensitivity band (-6 dB), bandpass filtered and backprojected to create thermoacoustic images in the plane of the ultrasound array. The same soundspeed and transducer array were used to estimate range and generate the ultrasound images onto which Bragg peak locations were overlaid. An air-gap phantom that displaced the Bragg peak by 6.5 mm demonstrated accuracy. Robustness to soundspeed errors was demonstrated in a waterbath as the assumed propagation speed scanner setting was altered by +/- 5%. Tissue-mimicking gelatin and a bone sample were introduced to demonstrate robustness to acoustic heterogeneity relative to ultrasound images of the underlying morphology. Results: Single ion pulse measurements sufficed during the helium run, but signal averaging was required for protons. Moreover, range and entry point into the target were estimated from data collected by transducers placed at least 6 cm distal to the Bragg peak. When ultrasound images depicted the air-target interface where the beam enters, estimates of the entry point agreed with ultrasound images and range estimates agreed with Monte Carlo simulations to within 300 mu m, even when thermoacoustic emissions traveled through a strongly scattering bone sample. Estimated Bragg peak locations were translated 6.5 mm by the air-gap phantom and correctly identified scenarios when the beam stopped inside the bone. Conclusions: Soundspeed errors dilate and acoustic heterogeneities deform ultrasound images. When thermoacoustic receivers are co-located with the ultrasound imaging array, the same transformations shift thermoacoustic range estimates. Hence, thermoacoustic range verification is robust relative to ultrasound images of underlying anatomy. When the treatment target is visible in ultrasound, e.g., prostate, online thermoacoustic range estimates could verify that the treatment spot is inside the target.

Research Organization:
Argonne National Laboratory (ANL), Argonne, IL (United States)
Sponsoring Organization:
USDOE Office of Science (SC), Nuclear Physics (NP)
Grant/Contract Number:
AC02-06CH11357
OSTI ID:
1503258
Alternate ID(s):
OSTI ID: 1483305
Journal Information:
Medical Physics, Vol. 46, Issue 1; ISSN 0094-2405
Publisher:
American Association of Physicists in MedicineCopyright Statement
Country of Publication:
United States
Language:
English
Citation Metrics:
Cited by: 16 works
Citation information provided by
Web of Science

References (22)

Experimental studies of the acoustic signature of proton beams traversing fluid media journal May 1979
Acoustic detection of high energy particle showers in water journal August 1979
Time resolved properties of acoustic pulses generated in water and in soft tissue by pulsed proton beam irradiation-A possibility of doses distribution monitoring in proton radiation therapy: Time resolved properties of acoustic pulses journal November 1991
Acoustic pulse generated in a patient during treatment by pulsed proton radiation beam journal January 1995
Submillimeter ionoacoustic range determination for protons in water at a clinical synchrocyclotron journal August 2017
Ionizing radiation‐induced acoustics for radiotherapy and diagnostic radiology applications journal May 2018
Ionoacoustic tomography of the proton Bragg peak in combination with ultrasound and optoacoustic imaging journal July 2016
Thermoacoustic range verification using a clinical ultrasound array provides perfectly co-registered overlay of the Bragg peak onto an ultrasound image journal July 2016
ATLAS with CARIBU: A Laboratory Portrait journal January 2016
Tissue-mimicking phantom materials for narrowband and ultrawideband microwave applications journal August 2005
SRIM – The stopping and range of ions in matter (2010)
  • Ziegler, James F.; Ziegler, M. D.; Biersack, J. P.
  • Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, Vol. 268, Issue 11-12 https://doi.org/10.1016/j.nimb.2010.02.091
journal June 2010
Thermoacoustic tomography—consistency conditions and the partial scan problem journal May 2004
Universal back-projection algorithm for photoacoustic computed tomography journal January 2005
Ex Vivo Thermoacoustic Imaging Over Large Fields of View With 108 MHz Irradiation journal August 2011
Two-stage ionoacoustic range verification leveraging Monte Carlo and acoustic simulations to stably account for tissue inhomogeneity and accelerator-specific time structure - A simulation study journal December 2017
Is Ultrasound Fusion a Reasonable Replacement for Computed Tomography in Guiding Abdominal Interventions? journal April 2016
Preliminary results on the feasibility of using ultrasound to monitor intrafractional motion during radiation therapy for pancreatic cancer: Feasibility of using ultrasound to monitor intrafractional motion journal August 2016
Fusion imaging of contrast-enhanced ultrasound and contrast-enhanced CT or MRI before radiofrequency ablation for liver cancers journal November 2016
A CT based correction method for speed of sound aberration for ultrasound based image guided radiotherapy: Speed of sound aberration correction in IGRT journal May 2011
On the significance of density-induced speed of sound variations on US-guided radiotherapy: SOS aberration correction between simulation and treatment journal September 2012
Feasibility of RACT for 3D dose measurement and range verification in a water phantom: Radiation acoustic computed tomography journal January 2015
How proton pulse characteristics influence protoacoustic determination of proton-beam range: simulation studies journal February 2016

Cited By (1)

Photoacoustic dose monitoring in clinical high-energy photon beams journal April 2019

Figures / Tables (10)